Contact an Account ManagerDetermination of PCBs and PAHs with PCB 28 / PCB 31 Separation in less than 10 min
Summary
Type: GC
Application #: 212920
Catalog#: 725850.30 - GC Column, Optima-XLB, 0.25mm, 0.25um, 30m
Detection: GC-MS:
Interface temperature: 280°C
Ion source temperature: 230°C
Quadrupol temperature: 150°C
Conditions: Injection volume: 1 µl, standard 0.005 ng/µl
Injection temperature: 250°C
Injection pulse: 1.38 bar in 1 min; splittless
Carrier gas: Helium, with purge flow of 60 ml/min
Temperature programm: 40°
Authors: Centre dAnalyses de Recherche
Source: Lab DHydrologie Illkirch (France) 2007
Details
Legend: Chromatogram B: 82% separation of PCB 28 / PCB 31 in less than 10 min
Chromatogram C: 75% separation of benzo[b]fluoranthene / benzo[k]fluoranthene in 17 min
1 naphthalene
2 2-methylnaphtalene
3